Practical Significance and Economic Efficiency of Transcutaneous Bilirubinometry

Aksenov D*
* Corresponding author
ISSN: 2640-2726  10.23880/pnboa-16000135  Received: January 17, 2019  Published: January 28, 2019
  views
 11 references
PDF
Keywords
Screening Transcutaneous bilirubin determination Jaundice Premature infants Infants
Abstract

Recently, a number of research papers have been published in the medical literature, demonstrating the effectiveness of the transcutaneous bilirubinometry method in both full - term and premature infants. This technique is widely implemented in most hospitals around the world and is a screening for pathological hyperbilirubinemia. The use of the device for transcutaneous bilirubinometry can significantly reduce the frequency of invasive studies of the level of bilirubinemia. The purpose of this analysis was to compare transcutaneous and invasive methods for determining bilirubin in the group of full-term and premature infants, as well as to determine the cost-effectiveness of this technique on the example of the JM-105. The device Jaundice Meter Model JM-105 (Draeger Medical Systems, Germany) uses two wavelengths and a double optical path system. The result of each definition is calculated as the average of the three dimensions. Resource charged battery 250 measurements. Lamp life 150000 measurements. No consumables. The following results were obtained in full-term children: in the range of bilirubin concentrations from 100 to 255 µmol/l between the results of serum bilirubin and transcutaneous measurement device JM-105" found a very close correlation - r=0.97, p<0.05. As a result of examination of premature newborns: at the level of bilirubinemia from 50 to 270 µmol / l, the correlation between the invasive method of determining serum ilirubin and transcutaneous bilirubinometry JM-105 (r=0.95). With an average cost of the device for transcutaneous bilirubinometry JM-105 about 8000 USD, which is the equivalent of about 8000 blood tests for bilirubin, taking into account the fact that this technique does not require additional consumables and significant investments for the entire period of the device, the use of this technique is economically feasible.

Purpose of Research

Was to compare percutaneous and invasive methods of bilirubin determination in full-term groups (37-40 weeks of gestation) and premature infants with gestational age (28-34 weeks), as well as to determine the cost-effectiveness of this technique on the example of JM- 105 [1].

Methods

The study included 120 newborns with a gestation period of 37-40 weeks [2, 3], with birth weight from 2500 to 4400 g and 44 premature infants with gestational age from 28 to 34 weeks, with birth weight in the range from 980 to 2720 grams[4]. Infants with documented cases of ABO(blood group system) or rhesus factor incompatibility, congenital malformations, hemoglobinopathy, and signs of liver damage were excluded from the study. In newborns aged 2 to 14 days of life, transcutaneous bilirubinometry was carried out by Drager JM-105 (produced by Drager Medical GmbH, Germany) [1]. In parallel, venous blood was taken for examination by liquid chromatography apparatus Dimension Rxl max (Siemens healthcare diagnostics, USA) in the laboratory. The transcutaneous measurement of bilirubin in the area of the forehead and chest of a premature newborn, covered from direct sunlight in the hair-free zone, and in the absence of hemorrhages, nevi or other skin abnormalities was carried out in the interval ± 30 minutes from the moment of determination of the total serum bilirubin (TBS). The use of transcutaneous bilirubinometer was carried out in accordance with the manufacturer's instructions [1].

All measurements in premature newborns were performed in the morning in natural light on the skin, forehead and chest. Bilirubin was measured again in children 24-72 hours after the primary measurement, depending on the dynamics of jaundice [2, 3, 4].

A total of 102 full-term infants underwent 281 parallel determination of bilirubin concentration by invasive and non-invasive methods, 44 premature infants underwent 70 parallel determination of bilirubin concentration by invasive and non-invasive methods. The level of bilirubinemia above 270 µmol/l was excluded from the study, as the sensitivity of the transcutaneous technique at such values decreased sharply. The calculation of the cost of performing one determination of blood bilirubin in the laboratory (excluding depreciation of equipment and the cost of electricity). To assess the economic efficiency of transcutaneous bilirubinometry, the JM-105 device is taken as an example. Statistical processing of the results was carried out using the application program Statistica (version 10.0).

Results

As a result of the examination of full-term and premature newborns, during the statistical analysis with the calculation of the pair correlation coefficient (r), the following results were obtained : in full-term newborns at the level of bilirubinemia from 100 to 255 µmol/l, a very high degree of correlation was observed between the indicators of total bilirubin capillary blood (TKB) measured by the traditional method and the indicators of transcutaneous determination of bilirubin (TcB) measured by the device "JM-105" (r=0.97, p<0.05). In premature newborns: the level of bilirubinemia from 50 to 270 µmol/l correlation between invasive method of determining TBS analyzer Dimension Rxl max and TcB on the fore head JM-105 (r=0,94), ТBS and TcB on the chest JM-105 (r=0.95). The analysis of the cost of one blood test for bilirubin, taking into account the reagent and consumables, in a particular hospital. The average cost of one blood test is 1 USD. With an average cost of the device for transcutaneous bilirubinometry JM-105 about 8000 USD, which is the equivalent of about 8000 blood tests for bilirubin, taking into account the fact that this technique does not require additional consumables and significant investments for the entire period of the device, the use of this technique is economically feasible[5, 6, 7, 8].

Conclusion

When screening for bilirubinemia in premature infants, the JM-105 demonstrated good efficacy in comparison with invasive methods for determining bilirubinemia. This technique can be used in routine practice of management of full-term and premature infants with hyperbilirubinemia [9]. Routine use of transcutaneous bilirubinometry not only reduces the frequency of invasive intervention in the newborn, but also will significantly reduce the cost of laboratory diagnosis of bilirubinemia in the hospital [10, 11].

Conflict of Interest

The authors of the study declare that there is no conflict of financial interests in connection with the preparation and conduct of this study, as well as the lack of any financial support for research.

References

  1. (2018) Drager Jaundice Meter Model JM-105 Instructions for Use. Drager pp: 1-3.
  2. Aksenov DV, Timofeeva LA, Degtyarev DN (2015) Comparison of the results of invasive and non- invasive methods of diagnosis of hyperbilirubinemia in newborns. Neonatology: news, opinions, training 2: 63-67.
  3. Volodin NN, Degtyarev D., Degtyarev AV, et al (2012) Tactics of conducting full-term and premature newborns with indirect hyperbilirubinemia. Neonatology: news, opinions, training 2: 113-132.
  4. Aksenov DV, Magnitskaya EA, Strogonov EV, Chusov KP, Kirtbaya AR, et al. (2017) A comparison of transcutaneous and invasive methods for the determination of bilirubinemia in premature infants. Neonatology: news, opinions, training 4: 79-87.
  5. Maisels MJ (2006) Historical perspectives: transcutaneous bilirubinometry. NeoReviews 6.: e217-e225.
  6. Raimondi F, Lama S, Landolfo F, Sellitto M, Borrelli AC, et al. (2012) Measuring transcutaneous bilirubin: a comparative analysis of three devices on a multiracial population. BMC Pediatrics 12: 70.
  7. Schmidt ET, Wheeler CA, Jackson GL, Engle WD (2009) Evaluation of transcutaneous bilirubinometry in preterm neonates. J Perinatol 29(8): 564-569.
  8. Bertini G, Pratesi S, Cosenza E, Dani C (2008) Transcutaneous bilirubin measurement: evaluation of Bilitest. Neonatology 93(2): 101-105.
  9. Keren R, Luan X, Friedman S, Saddlemire S, Cnaan A, et al. (2008) A comparison of alternative risk assessment strategies for predicting significant neonatal hyperbilirubinemia in term and near-term infants. Pediatrics 121(1): e170-e179.
  10. Nagar G, Vandermeer B, Campbell S, Kumar M (2013) Reliability of transcutaneous bilirubin devices in preterm infants: a systematic review. Pediatrics 132(5): 871-881.
  11. Rubio A, Epiard C, Gebus M, Deiber M, Samperiz S, et al. (2017) Diagnosis Accuracy of Transcutaneous Bilirubinometry in Very Preterm Newborns. Neonatology 111(1): 1-7.

Cite this article

BibTeX
APA
RIS
@article{aksenov2019,
  title   = {Practical Significance and Economic Efficiency of Transcutaneous Bilirubinometry},
  author  = {Aksenov D},
  journal = {Pediatrics & Neonatal Biology Open Access},
  year    = {2019},
  volume  = {4},
  number  = {1},
  doi     = {10.23880/pnboa-16000135}
}
Aksenov D (2019). Practical Significance and Economic Efficiency of Transcutaneous Bilirubinometry. Pediatrics & Neonatal Biology Open Access, 4(1). https://doi.org/10.23880/pnboa-16000135
TY  - JOUR
TI  - Practical Significance and Economic Efficiency of Transcutaneous Bilirubinometry
AU  - Aksenov D
JO  - Pediatrics & Neonatal Biology Open Access
PY  - 2019
VL  - 4
IS  - 1
DO  - 10.23880/pnboa-16000135
ER  -