International Journal of Nuclear Medicine & Radioactive Substances (IJNMRS)

ISSN: 2689-8020

Research Article

Gamma Evaluation with Octavius 4D Phantom for PreTreatment of Modern Radiotherapy Treatment Techniques

Authors: Srivastava RP*, Basta K, Thevissen K, Junius S, Vandeputte K and Wagter CD

DOI: 10.23880/ijnmrs-16000117

Abstract

Purpose: In modern radiotherapy techniques such as intensity-modulated radiation therapy (IMRT) and volume modulated arc therapy (VMAT), the quality assurance (QA) process is vital. The goal of the study was to verify the treatment planning dose delivered during delivery of complex treatment plans. The QA standard is to perform patientspecific comparisons between planned doses and doses measured in a phantom. Materials and Methods: Ninety-five complex IMRT and VMAT plans for different pathologies planned using Eclipse treatment planning system (TPS). The Octavius 4D phantom has been used to verify patient specific quality assurance of all VMAT plans calculating with different algorithms. Conclusion: Overall, good agreement was observed between measured and calculated doses in most cases with gamma values above 1 in >95% of measured points in volumetric 3D analysis and 1 in >90% in 2D analysis. The Octavius 4D phantom is an effective and efficient method for patient specific QA.

Keywords: IMRT; VMAT; Quality Assurance; Gamma Value; Treatment Planning System