Role of MDCT Coronary Angiography with 64-Slice Technology to Detect the Coronary Artery Diseases
Aims: The aim of our study was to investigate the accuracy of 64-slice computed tomography (CT) for assessing hemodynamically significant stenosis of coronary arteries. Methods and Results: CT angiography was performed in 62 patients (50 male, 12 female) with suspected coronary artery disease and compared with invasive coronary angiography. All vessels were considered for the assessment of significant coronary artery stenosis. Sixty two patients were identified as having significant coronary stenosis on invasive angiography with 56% (140/248) affected segments. CT correctly identified all 7 patients having no significant stenosis on invasive angiography. Overall sensitivity for classifying stenosis was 96%, specificity was 80%, positive predictive value was 98%, and negative predictive value was 67%. Conclusion: Sixty-four-slice CT provides a high diagnostic accuracy in assessing coronary artery stenosis.
Introduction
Advances in multidetector CT (MDCT) technology with sub millimeter slice collimation (0.5 mm) and high temporal resolution ((330 ms) permit contrast-enhanced imaging of coronary arteries and coronary plaque during a single breath hold [1]. Appropriate patient preparation, detailed technical and technological knowledge with regard to recognition of typical imaging artifacts (such as beam hardening or motion artifacts), and the adequate choice of post processing techniques to detect stenosis and plaque are prerequisites to achieving diagnostic image quality. A growing number of studies have suggested that 64-slice coronary CT angiography is highly accurate for the exclusion of significant coronary artery stenosis (>50% luminal narrowing), with negative predictive values of 97%–100%, in comparison with invasive selective coronary angiography. In addition, several studies have indicated that MDCT also can detect calcified and no calcified coronary atherosclerotic plaques [2], especially in proximal vessel segments, showing a good correlation with intracoronary ultrasound. Studies on clinical utility, cost, and cost-effectiveness are now warranted to demonstrate whether and how this technique can change and improve the current management of patients with suspected or confirmed coronary artery disease [3]. Over the last 10 years a dramatic improvement in multi detector CT (MDCT) technology has occurred. The ability to noninvasively image the coronary artery lumen and wall and obtain information on the presence, severity, and characteristics of coronary artery disease (CAD), including the visualization of luminal obstruction and atherosclerotic plaque, constitutes an attractive addition to currently available diagnostic tools, such as nuclear perfusion imaging or invasive selective coronary angiography, for the work-up of patients with known or suspected CAD [3].
Established Modalities in Cardiac Imaging
Various invasive and non-invasive imaging techniques are in use for cardiac diagnosis. The potential role of CT imaging as a new modality in cardiac disease management derives from the limitations in clinical application and the shortcomings in economic efficacy of the techniques that are available today. a) Invasive Imaging (Conventional angiography) b) Noninvasive imaging
- Echocardiography
- Cardiac Perfusion Imaging
- Magnetic resonance Imaging
- CT angiography
Prevalence
25% of all deaths in the USA are related to CAD (Jones and Eaton). About 50% of these patients die after acute myocardial infarction without prior symptoms. The main risk factors to develop CAD are age, sex (men are at higher risk than women), high blood pressure, high blood lipid level, smoking, diabetes, overweight, family history, lack of exercise and mental stress. Approx. 1 in 226 or 0.44% or 1.2 million people in USA (source statistic for calculation: estimated 1.2 million new or recurrent coronary attacks in the USA 2004 (American Heart Association).
Incidence Extrapolations for USA for Coronary Heart Disease
1,200,000 per year, 100,000 per month, 23,076 per week, 3,287 per day, 136 per hour, 2 per minute, 0 per second. [Source statistic for calculation: "estimated 1.2 million new or recurrent coronary attacks in the USA 2004 (American Heart Association).
Incidence Extrapolations for USA for Coronary Heart Disease
1,200,000 per year, 100,000 per month, 23,076 per week, 3,287 per day, 136 per hour, 2 per minute, 0 per second. [Source statistic for calculation: "estimated 1.2 million new or recurrent coronary attacks in the USA 2004 (American Heart Association).
Epedimiology
The main risk factors to develop CAD are Age, Sex (men are at higher risk than women), High blood pressure, High blood lipid level, Smoking, Diabetes, Overweight, Family history, Lack of exercise and mental stress (Assmann et al. 1999). CAD is the symptomatic presentation of coronary atherosclerosis, which causes narrowing of the coronary vessel lumen due to lipid-rich, fibrous or calcified plaques that adhere to the vascular walls. The reduction in blood and oxygen supply causes ischemia in the myocardium. The patient perceives this as an attack of chest pain (angina pectoris) and, in advanced stages, as acute myocardial infarction with necrotic changes of the heart muscle if coronary vessels are severely obstructed or totally occluded.
Instrumentation
Aquilion 64-Slice MDCT, Toshiba, Japan • 64-slice scanner has increased slices per gantry rotation (64 slices) and faster gantry speed (330 ms/rotation), which translate into superior spatial resolution (0.5) and temporal resolution (165 vs. 188 ms). The reduction in voxel size makes the distinction between hypo intense soft plaque and blood pool contrast more evident (Figure 1).

Objective
The study is designed with following objectives: a) To measure the Influence of calcium score & grading of stenosis on CT Angio and conventional angio. b) To measure the accuracy of 64-slice MDCT by comparing the Cardiac CTA and Conventional Coronary Angiography. c) To detect the coronary artery disease through CT Angio & its comparison with conventional coronary angiography.
Rational
This study may help: a) Most of our persons in our community are hypertensive, obese, family History, lack of exercise, Smoker, high Cholesterol level, diabetic etc. b) Patients and coronary heart disease is main a cause of narrowing or blockage of vessels or causing cardio vascular diseases. c) CT angiography is initial tool for imaging or screening the cardio vascular diseases (atherosclerosis, stenosis) in those patients. d) In CT Angiography we can determine the nature of plaque and grading of stenosis can be done. This is non- invasive procedure; fast imaging done in CT angiography and it has high spatial resolution and minimum slice thickness required. e) Although in angiography (conventional angiography) is gold standard procedure for determining the nature of plaque but for this intra-vascular ultrasound is required which is too much expensive and require arterial puncture (invasive procedure). So CT is tool for initial imaging the location of plaque and determines the nature of plaque and grading of stenosis. f) Reduce lack of agreement between imaging findings and intraoperative findings.
Material and Methods
Subject & Study Design
This is a non-randomized, observational study design conducted at Heart, Brain & Body Scan, and Diagnostic Centre of radiology and angiography department of Doctor S’ Hospital Lahore.
Study Population
In 62 consecutive patients scheduled to have invasive coronary angiography for suspected coronary artery disease, MSCT was also performed in same patients. The human investigation committee approved the study protocol, and all patients gave informed consent.
Inclusion Criteria
a) Suspected coronary artery disease. b) Differential diagnosis of symptomatic patients with atypical chest pain. c) Risk stratification in asymptomatic individuals with intermediate to high risk for future myocardial event. d) Follow up of post stenting or post-CABG patients [4].
Exclusion Criteria
Exclusion criteria included irregular heart rate, patients at risk for iodinated contrast agents (congestive heart failure, dye allergy, or elevated serum creatinine >1.5 mg/dl), or contraindications to beta- blocking drugs [4].
Patient Preparation
Patients not already on beta-blocking drugs received 100 mg of atenolol for heart rates over 65 beats/min, or 50 mg of atenolol for heart rates under 65 beats/min but over 50 beats/min, 1 h before MSCT imaging. Heart rate, electrocardiogram (ECG), and blood pressure were monitored, and additional intravenous metoprolol (5 to 30 mg) was administered to achieve a target heart rate <65 beats/min [5]. However, no patient was excluded because of a heart rate above this target. Sublingual nitroglycerin 0.4 mg was given 1 min before image acquisition [3].
Scan Protocol and Image Reconstruction
All patients were scanned on a 62-slice MSCT scanner (Aquilion 64-Slice MDCT, Toshiba, Japan). An initial nonenhanced ECG-gated scan was performed for calcium scoring [2]. CT scans performed on a 64-slice scanner with a 0.37 s rotation time A bolus of 80 mL iodinated contrast media (Visipaque, Ultravist, omnipaque 320 mg/mL) was injected into an antecubital vein at a flow rate of 5 mL/s, followed by a 50 mL saline chasing bolus. Start delay was defined by bolus tracking in the ascending aorta and scan start was automatically initiated 5 s after reaching the threshold [140 HU (HU, Hounsfield Units)]. After this, scanning was performed from the tracheal bifurcation to the diaphragm using the following parameters: X-ray tube potential 120 kV, effective tube current 680 mA, slice collimation 64x0.5 mm, table feed 9.2 mm/rotation, and pitch 0.24. All CT scans were performed using an implemented fully automated real- time anatomy based dose regulation. The overall scan time was shorter than 12 s (median 11.2 s, range 10.7– 11.9 s) and the mean total time for the examination was <15 min. We used retrospective electrocardiographic (ECG) gating for optimal heart phase selection [6]. Estimated effective radiation dose was 13 mSv for men and 18 mSv for women. Electrocardiographically gated datasets were reconstructed automatically at 65% of the R-R cycle length and 35% of the R-R cycle length to approximate end-systole and end-diastole, respectively. Additional reconstruction windows were constructed after examination of datasets in motion artifacts was present [4] (Figure 2).

Statistical Analysis
MDCT Angiographic Analyses
Calcium scores in Agatston U were analyzed using CACS software (Aquilion 64-Slice MDCT, Toshiba, Japan); MSCT angiograms were analyzed on a three-dimensional workstation (VITREA). Scans were analyzed by consensus of two observers unaware of the clinical data and blinded to results of the invasive quantitative coronary angiography (QCA). Each lesion identified was examined using MIP (maximum intensity) and MPR (multiplanar reconstruction) VRT, techniques along multiple longitudinal axes and transversely. Lesions were classified by the maximal luminal diameter stenosis seen in any plan. Lesions that could not be quantified to within 25% diameter stenosis because of hazy or irregular borders were classified by a qualitative severity scale: 0 = no stenosis, 1 = 1% to 25% stenosis, 2 = 26% to 50%, and 3 = 51% to >75%, = 4 [4].
Invasive Angiographic Analyses
Invasive coronary angiograms were evaluated by a single observer blinded to the MSCT results. Segmental disease was analyzed in each vessel using the same 15- segment model employed for MSCT analysis. Stenosis severity in each segment was classified according to the maximal luminal diameter stenosis present in each segment. Lesions were examined in orthogonal views, and stenosis severity determined.
Comparative Analyses
Quantitative per-lesion analysis compared the maximal percent diameter stenosis of the most severe lesion in each segment by each imaging modality. The accuracy of MSCT to detect significant disease is compared to invasive QCA according to the following analyses: 1) per-artery analysis, comparing each artery, examining the presence of significant lesions in each of the major coronary vessels (right coronary artery, left circumflex, left anterior descending, and left main) [7]. Per-patient analysis evaluating the presence of any significant lesion in a given patients. In determining MSCT lesion severity, quantitative values were always used if available. For analysis of sensitivity, specificity, and positive and negative predicative accuracy, QCA lesion severity was used as the “gold standard.
Results
Descriptive Statistical Analysis
a) Frequency Distribution (Figure 3)

b) Calcium Scoring on CT scan (Figure 4)

Influence of Calcium Score on MSCT Accuracy
We examined the accuracy of MSCT to detect a significant stenosis in a given patient according to calcium score (Agatston U). The Effect of Coronary Calcification on Diagnostic Accuracy of Coronary MSCT Compared to QCA (Figures 5 & 6, Table 1).


| Patients’ calcium score * | Sensitivity | Specificity | PPV | NPV | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 0–100 (n = 37) | 11/13 (85%) | 12/14 (86%) | 11/13 (85%) | 12/14 (86%) | ||||||||
| >300 (n = 4) | 4/4 (100%) | 12/14 (86%) | 4/6 (67%) | 12/12 (100%) |
Table 1: Showing the accuracy of 64 slice CT for Calcium scoring patients.
a) Ho: CTCA = CCA b) HI: CTCA =/= CCA c) Level of significance: 95%, Alpha = 0.05 d) Test Statistics: Spearmen Correlation = r e) MDCT compared to CCA for quantification of lesion severity (Table 2). Calculations:
| Correlations | |||
|---|---|---|---|
| Conventional Coronary Angiography | Cardiac CT Angiography | ||
| Pearson Correlation | 1 | .973(**) | |
| Sig. (2-tailed) | .000 | ||
| Pearson Correlation | .973(**) | 1 | |
| Sig. (2-tailed) | .000 | ||
| ** Correlation is significant at the 0.01 level (2-tailed) |
Table 2: The spearman correlation coefficient between the two modalities CTCA and CCA is 0.973 (p<0.05).
f) Critical region if P<α (0.000<0.05) then we reject the null hypothesis g) Since we calculate that spearman correlation and hence we conclude that reject Ho & accept alternative hypothesis because P<α (0.000<0.05)s
MSCT Compared to QCA for Detection of Significant Stenosis
Diagnostic Accuracy of Coronary MSCT Compared to QCA for Detection of Lesions (Table 3)
| Sensitivity | Specificity | PPV | NPV | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Patients (n = 62) | 55/57 (96%) | 4/5 (80%) | 55/56 (98%) | 4/6 (67%) | ||||||||
| Arteries (n = 248) | 140/152 (92%) | 90/96 (94%) | 140/146 (96%) | 90/102 (88%) |
Table 3: A total of 248 arteries could be analyzed. On a per-artery basis, MSCT had a sensitivity of 92% (140 of 152) and a speci
All 62 patients could be analyzed for the presence of disease. Multislice CT correctly detected significant disease in 55 of 57 cases for an overall sensitivity per- patient of 96%. Multislice CT documented absence of significant disease in 4 of 5 patients for an overall specificity per-patient of 80%. The positive predictive value for any significant disease was 98% (55 of 56), and the negative predictive value was 67% (4 of 6). Accurate determination of the presence or absence of significant coronary disease was made in 59 of 62 patients (95%) (Figures 7-14).








Discussion
Observations from this study demonstrate that this 64-slice MSCT scanner consistently provides high-quality noninvasive coronary arteriograms that accurately delineate the presence or absence of significant lesions within the entire coronary tree in a broad spectrum of patients, including those with marked coronary calcification, relatively high heart rates, and obesity. In aggregate, these studies document that noninvasive coronary MSCT imaging can accurately determines the presence or absence of significant coronary lesions. However, prior studies, with one exception employed qualitative analyses only. Furthermore, in contrast to the present study, these prior studies excluded many “real- world” patients because of high heart rates, coronary calcification, or obesity, in addition to excluding all vessels <1.5 mm in diameter. In comparison to 16-slice MSCT scanners, the present 64-slice scanner has increased slices per gantry rotation (64 vs. 16) and faster gantry speed (330 ms/rotation vs. 375), which translate into superior spatial resolution (0.4 vs. 0.5 mm) and temporal resolution (165 vs. 188 ms). The reduction in voxel size makes the distinction between hypointense soft plaque and blood pool contrast more evident. Smaller voxel size also reduces partial volume effects, minimizing the degree of calcium blooming and beamhardening artifacts. Partial volume effects are due to averaging of different densities within a single voxel. Calcium deposits are metal density and thus overwhelm the density of other tissues in the same voxel. Beam hardening is due to the attenuation of low-energy X-ray by very dense structures such as calcium. A higher energy beam, causing a darker appearance that can be mistaken for plaque, therefore penetrates adjacent pixels. All these effects can be modified, but not eliminated, by the smaller voxel size produced by the 64-slice scanner. The efficacy of this scanner in ameliorating imaging difficulties is shown in an overall sensitivity of 95% and specificity of 90% for the detection of angiographically significant stenoses even in the presence of high coronary calcium scores [1].
Invasive Imaging (Conventional angiography)
Selective coronary angiography (CA) is considered the gold-standard for diagnosis of the coronary artery lumen [8]. A catheter is inserted into the arterial system via a peripheral artery (Judkins technique) and moved through the aorta into the origins of the coronary arteries. The distribution of contrast agent, administered via the catheter into the coronary arteries, is imaged in several views with planar X-ray technique in order to identify narrowing of the vessel lumen. The maximum 2-di- mensional resolution that can be achieved is about 0.15 mm. Basic cardiac function parameters (e.g. ejection fraction) can be derived after administration of contrast agent into the ventricle and assessment of the ventricle size in diastole and systole. For patients with suspected CAD, the diagnostic procedure can be combined with PTCA or stent insertion in the same session. Although evaluation of the vessel wall and of coronary plaques is not possible, CA is currently the standard technique to detect and evaluate coronary artery stenosis and provides a basis for decisions for further work-up. However, due to its invasiveness with possible complications (Harrison) and the considerable amount of applied X-ray dose (3-6 mSv on average) (Leung and Martin) there is great interest in noninvasive imaging with less X-ray radiation and less direct interaction of the physician.
Evaluation of the vessel wall and of atherosclerotic plaques has become feasible with the advent of intravascular ultrasound (IVUS). IVUS applied to the coronary arteries allows for the assessment of the lumen diameter and plaque dimension. Plaque morphology can be evaluated in terms of lipid content and identification of fibrous and calcified material. However, IVUS has not yet been established as a routine clinical tool because of its high invasiveness, restriction to proximal coronary segments and high costs [6].
Non-invasive Imaging
The non-invasive cardiac imaging tools that are routinely used today are echocardiography, nuclear imaging and magnetic resonance imaging.
Echocardiography (cardiac ultrasound) allows assessment of shape and thickness of the cardiac walls and of the heart valves. Cardiac function can be evaluated via wall motion, and cardiac blood flow assessment is possible with Doppler sonography techniques. Atherosclerotic disease can be assessed in the carotid arteries and in the great thoracic vessels. However, imaging of the coronary arteries and detection of coronary plaques and stenosis are not yet possible.
Nuclear imaging can be used for evaluation of cardiac function, myocardial perfusion and myocardial viability. Magnetic resonance imaging MRI): Dedicated scan sequences have been developed for ECG-correlated imaging with high temporal resolution down to 20-50 ms that allow for true 4D imaging of the heart.
Cardiac imaging with MDCT
a) CT scans performed on a 64-slice scanner with a 0.37 s rotation time (Aquilion 64-Slice MDCT, Toshiba, Japan). A bolus of 80 mL iodinated contrast media (Visipaque, Ultravist, omnipaque 370 mg/mL) was injected into an antecubital vein at a flow rate of 5 mL/s, followed by a 50 mL saline chasing bolus. Start delay was defined by bolus tracking in the ascending aorta and scan start was automatically initiated 5 s after reaching the threshold [140 HU (HU, Hounsfield Units)]. After this, scanning was performed from the tracheal bifurcation to the diaphragm using the following parameters: X-ray tube potential 120 kV, effective tube current 680 mA, slice collimation 64x0.5 mm, table feed 9.2 mm/rotation, and pitch 0.24. All CT scans were performed using an implemented fully automated real- time anatomy based dose regulation. The overall scan time was shorter than 12 s (median 11.2 s, range 10.7– 11.9 s) and the mean total time for the examination was <15 min. We used retrospective electrocardiographic (ECG) gating for optimal heart phase selection. b) The implemented adaptive cardio volume approach was used for data reconstruction. Depending on the heart rate throughout the examination, axial slices were reconstructed synchronized to the ECG by a single sector (65 b.p.m) or two sector algorithm (>65 b.p.m) using data from one or two consecutive heartbeats [2]. When necessary, R-wave indicators were manually repositioned to improve the quality of synchronization. Images were reconstructed at 10% intervals of the cardiac cycle to allow assessment of coronary arteries at that cardiac phase with minimal vessel motion. Slices with a thickness of 0.75 mm (increment 0.5 mm) and a medium soft-tissue reconstruction kernel were used for evaluating coronary arteries [6].
Conclusion
My results indicate high quantitative and qualitative diagnostic accuracy of 64-slice MSCT in comparison to QCA in a broad spectrum of patients. The sensitivity for minimal to mild coronary artery diseases is greater of Cardiac CTA as compared to CCA, because MDCT consists of 0.5 collimated slice width and temporal resolution is less than 0.33 sec, so spatial resolution is also greater. MDCT has high efficiency to pick very minute calcium plaques in coronary artery vessels. MDCT can also differentiate between soft, fibrous and hard or calcium plaques with the help of different Hounsfield units but CCA cannot differentiate between these plaques.
References
-
Raff GL, Gallagher MJ, O’Neil WW, Goldstein JA (2005) Diagnostic Accuracy of Non-Invasive Coronary Angiography using 64-slice Spiral Computed Tomography. J American College of Cardiology 46(3): 552-557.
-
Hoffman MHK, Shi H, Schmid FT, Gelman H, Brambs HJ, et al. (2004)Non-invasive Coronary Imaging with MDCT in Comparison to Invasive Conventional Coronary Angiography: A Fast Developing Technology. American J Roentgenology 182(3): 601- 608.
-
Hoffman U, Ferencik M, Cury RC, Pena AJ (2006) Coronary CT Angiography. J Nucl Med 47(5): 797- 806.
-
Leschka S, Alkadhi H, Plass A, Desbiolles L, Grünenfelder J, et al. (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26(15): 1482-1487.
-
Ropers D, Baum U, Poehl K, Anders K, Ulzheimer S, et al. (2003) Detection of coronary artery stenoses with thin-slice multi-detector row spiral computed tomography and multiplanar reconstruction. Circulation 107(5): 664-666.
-
Fine JJ, Hopkins CB, Ruff N, Newton FC (2005) 64- slice Cardiovascular CTA, a clinical accuracy evaluation. Department of Investigator Initiated Research, South Carolina Heart Centre, Columbia, South Africa.
-
Nieman K, Cademartiri F, Lemos PA, Raaijmakers R, Pattynama PMT, et al. (2002) Reliable noninvasive coronary angiography with fast submillimeter multislice spiral computed tomography. Circulation 106(16): 2051-2054.
-
Bedi HS, Accuracy of 64-slice CT Coronary Angiography. Christian Medical College and Hospital, Ludhiania, India.
- Ultrasound Guided Therapeutic Nerve Blocks
- Cyclops Lesion Without ACL Reconstruction: A Rare Case in a Patient with Intact Anterior Cruciate Ligament and Tibial Plateau Fracture
- Dosimetric Comparison between Two Dose Calculation Algorithms in SBRT Treatment of Lung Cancer in Ring-based and C-arm Radiation Therapy Equipment
- Adolescent Testicular Adrenal Rest Tumors: A Case Report and Review of the Literature
- Giant Intrathoracic Lipoma: A Rare Presentation
- Image of a Right Renal Angiomyolipoma Complicated by Hemorrhage